Vapor Flow Calibration
Use to calibrate the type of vapor to be used when using the optional vapor generator. Associating the analysis data with a vapor generator calibration file allows the conversion of peak area data to volume data. The calibration can be performed either before or after performing the analysis. A single calibration file can be used with multiple analysis files.
- Go to File > Open > Sample information.
- Enter a calibration file name (.CAL).
- Click OK, then click Yes to confirm.
- Complete the remaining fields on the Sample Description window, as needed.
| If the window displays in Basic presentation, click Advanced in the view selector drop-down at the bottom of the window. |
- Select the Analysis Conditions tab.
- Select AutoChem III in the View conditions for drop-down list.
- Click Insert, then select Vapor Flow Calibration.
-
Complete the dialog box according to this table:
| Selections | Description |
|---|---|
| Carrier Gas/Loop Gas [drop‑down box] | Select the gases to be used for this step and enter the Flow rate. |
| Description [text box] | Description for this calibration. |
| Flask Temperatures [group box] | Enter a value between 20.0 and 120.0 degrees for the Minimum and Maximum. |
| Filament temperature [selection] | Enter the temperature for the filament. |
| Number of steps [text box] | Enter the number of steps (between 3 and 7) for this calibration. |
| Reflux Temperatures [group box] | Enter a value between 20.0 and 100.0 degrees for the Minimum and Maximum. |
| Set external trigger [check box] | If selected, the contact closure used to trigger an external mass spectrometer will be activated. If deselected, the contact closure will be deselected. |
| Temperature [group box] | Enter a value between 20.0 and 150.0 degrees for the Valves and Detector block. |
| Vapor [drop‑down box] | Select the vapor to be used. |
- Click OK.
- Go to Unit [n] > Sample Analysis.
- Click Browse and select the sample file just created. Make edits as required.
- Click Start.
- Select an option from the View drop-down list to observe the progress of the analysis.
- When the analysis completes, close the Analysis window.
- When the displayed sample temperature reaches the ambient temperature, open the furnace. Use gloves and remove the sample tube. These fields do not require editing; the Reflux and Flask temperatures entered on the Vapor Calibration Experiment window are used.
- To edit the calibration file, go to Options > Signal Calibration > New.
|
Use the cotton gloves provided in the accessory kit when handling heated surfaces. These cotton gloves are not intended to protect hands when heated surfaces are above 60 °C. |
- Select Load Calibration Data and select the sample file for this calibration. The data contained in the file are automatically loaded.
- Access the Calculations document on the Micromeritics web page (www.micromeritics.com) for information on calculating the volume of each vapor injection, then enter the values for each peak.
- Assess the plot, Goodness of Fit, and Coefficients to decide if the calibration file is acceptable. Use your laboratory’s standards to determine what level of linearity is acceptable. As a general guideline, use a calibration file with a very low value for Goodness of Fit (less than 1 to 2% of the signal), when 1st or 2nd Degree is specified.
- To delete any peaks that are outliers, highlight the peak data in the table and click Delete Peak.
- Evaluate the results when the degree is changed (select a different degree from the drop-down list).
- When satisfied with the calibration file, click Save then click Close.





