Loop Volume

It is preferred for the sample to require at least two doses of gas, but no more than ten doses, before the reaction ends — depending on the environment. Some factors that influence the number of doses required are sample size, the active surface area per unit of sample, and the size of the loop. If the pulse chemisorption analysis requires more doses or fewer doses, change either the sample size or the loop size.

Three loops of different sizes are provided with the analyzer — each loop must be calibrated prior to use. As an alternative to replacing the loop, the loop volume may be changed by adjusting its temperature. If the loop volume is changed by changing its temperature, results will be more accurate if the loop volume is recalibrated at the new temperature.

Loop Calibration

Before data from an analysis that uses an injection loop can be reduced, a loop calibration experiment must be performed. An independent loop calibration can be performed either before or after the analysis.

The following example assumes that the loop calibration will be performed after the analysis. A Loop Calibration step can be included in a pulse chemisorption analysis.

Preparation

Pretreatment Degas by flowing inert gas (such as helium, argon, or nitrogen) over the sample while ramping the temperature.
Analysis Pulse the loop gas over the sample until the peak area remains constant.
Trap No trap is needed during pretreatment. During the analysis, use the trap fixture with a Dewar of water in place as a short delay path. This delay compensates for the flow disturbance caused by each loop injection.
Pressure regulator Gas cylinders should be set to a level between 14 and 16 psig (95 and 110 kPag).
Furnace temperature Select a temperature high enough to remove any contaminants or moisture, but not so high as to cause sintering or fusing of the sample. Ensure the Done step is set to return the sample temperature to ambient.