Pulse Chemisorption Analysis Tutorial
A Pulse chemisorption analysis determines active surface area, percent metal dispersion, and active particle size by applying measured pulses of reactant gas to the sample, depending upon what technique is used. Pulse chemisorption is performed in the same manner as TPD, except that the sample is dosed with the analysis gas using the injection loop until each active site has reacted. After the active sites have completely reacted, the discretely injected gas volumes emerge unchanged. If TPD is performed after Pulse chemisorption, additional information about the distribution of active sites and the strength of active sites is collected.
The amount chemisorbed is the difference between the total amount of reactant gas injected and the amount that did not react with the active sites of the sample. The size of each pulse is determined by the loop, which is located on an electrically operated valve.
Procedure
| In the following example, Pulse chemisorption is performed on a sample of Pt/Al2O3 with CO. Alter the analysis conditions to accommodate a different analysis. |
| Before performing an analysis, ensure the sample and analyzer are adequately prepared. |
- Obtain the sample mass, then install the loaded sample tube and thermocouple on the analyzer. Close the furnace around the sample tube.
- Create a sample file containing the appropriate analysis conditions and report options.
- Go to File > New Sample.
- Complete the Sample Description window using appropriate values.
- Select the Analysis Conditions tab.
- Select AutoChem III in the View conditions for drop-down box.
- Insert the following experiment steps. After each step, click OK to close each window to move to the next step.
Experiment Steps:
- Insert > Experiment
| Description | Enter a description of the experiment |
| Type of Analysis | Select Pulse |
- Insert > Instrument Settings
| Carrier/Reference Gas | Loop, Injection, or Blend Gas | Preparation Gas | ||||||
| None | None | Helium | ||||||
|
|
|
||||||
|
||||||||
|
||||||||
| Vapor valve | Blend valve | Loop valve | Analysis valve | Trap valve | ||||||||||
|
|
|
|
|
||||||||||
|
|
|
|
|
| Vapor | Valves | Sample port | Sample | Rate | ||||||||||||
|
|
|
|
|
||||||||||||
|
|
|||||||||||||||
|
|
|||||||||||||||
| Block temperature | 100 | °C | ||
| Enable detector | ||||
|
245 | °C |
- Insert > Wait
|
Wait 5 minutes. |
- Insert > Instrument Settings
| Carrier/Reference Gas | Loop, Injection, or Blend Gas | Preparation Gas | ||||||
| None | None | Hydrogen | ||||||
|
|
|
||||||
|
||||||||
|
||||||||
| Vapor valve | Blend valve | Loop valve | Analysis valve | Trap valve | ||||||||||
|
|
|
|
|
||||||||||
|
|
|
|
|
| Vapor | Valves | Sample port | Sample | Rate | ||||||||||||
|
|
|
|
|
||||||||||||
|
|
|||||||||||||||
|
|
|||||||||||||||
| Block temperature | 100 | °C | ||
| Enable detector | ||||
|
245 | °C |
- Insert > Temperature Ramp
| Sample ramp |
|
- Insert > Instrument Settings
| Carrier/Reference Gas | Loop, Injection, or Blend Gas | Preparation Gas | ||||||
| Helium | None | Helium | ||||||
|
|
|
||||||
|
||||||||
|
||||||||
| Vapor valve | Blend valve | Loop valve | Analysis valve | Trap valve | ||||||||||
|
|
|
|
|
||||||||||
|
|
|
|
|
| Vapor | Valves | Sample port | Sample | Rate | ||||||||||||
|
|
|
|
|
||||||||||||
|
|
|||||||||||||||
|
|
|||||||||||||||
| Block temperature | 100 | °C | ||
| Enable detector | ||||
|
245 | °C |
- Insert > Wait
|
Wait 30 minutes. |
- Insert > Temperature Ramp
| Sample ramp |
|
- Insert > Instrument Settings
| Carrier/Reference Gas | Loop, Injection, or Blend Gas | Preparation Gas | ||||||
| Helium | CO/He | None | ||||||
|
|
|
||||||
|
||||||||
|
||||||||
| Vapor valve | Blend valve | Loop valve | Analysis valve | Trap valve | ||||||||||
|
|
|
|
|
||||||||||
|
|
|
|
|
| Vapor | Valves | Sample port | Sample | Rate | ||||||||||||
|
|
|
|
|
||||||||||||
|
|
|||||||||||||||
|
|
|||||||||||||||
| Block temperature | 100 | °C | ||
| | Enable detector | |||
|
245 | °C |
-
Insert > Wait
|
Wait 5 minutes. |
- Insert > Wait
|
Wait until baseline and temperature are stable. |
- Insert > Wait
|
Wait 5 minutes. |
- Insert > Start Recording
|
One measurement every 0.1 s |
- Insert > Start Repeat
|
Repeat until peaks are equal or 20 times. |
- Insert > Dose
Wait for 2 minutes.
|
- Termination: Double click Termination step in the experiment list box (or click Termination, then click Edit).
| Vapor | Valves | Sample port | Sample | Rate | ||||||||||||
|
|
|
|
|
||||||||||||
|
|
|||||||||||||||
|
|
|||||||||||||||
| Block temperature | 100 | °C | ||
| Enable detector | ||||
|
245 | °C |
- Select the Report Options tab and modify values as needed.
- Click Save, then click Close.
- Start the analysis.
- Go to Unit > Sample Analysis. From the Files list box, select the sample file created in the previous step. Edit the file as needed. Click Next.
- From the drop-down list, select the calibrations associated with each experiment in the sample file (if any). For this example, select None. Click Next.
| Calibration files can also be associated with a sample file after analysis using Calibration button in the Peak Editor. |
- Read the cautionary window and make any necessary changes.
- Click Start to start the analysis.
When the analysis ends, the furnace begins to lower the sample to room temperature. To speed cooling of the sample, use the optional CryoCooler or AutoCool.
|
Use the cotton gloves provided in the accessory kit when handling heated surfaces. These cotton gloves are not intended to protect hands when heated surfaces are above 60 °C. |
When the displayed sample temperature reaches the ambient temperature, open the furnace. Using gloves, remove the sample tube.





