Show Instrument Schematic
Unit [n] > Enable Manual Control
Unit [n] > Show Instrument Schematic
|
Standard |
|
Trap Heater |
|
CryoCooler |
|
Vapor |
| Valve | Description |
|---|---|
| 1 through 6 | Reference and carrier gas inlet ports |
| 11 through 16 | Loop gas inlet ports |
| 17 through 26 | Preparation gas inlet ports |
| Icon or Symbol | Description | |||
|---|---|---|---|---|
|
AutoCool. Displays only if a CyroCooler is not installed. | |||
|
|
Auxiliary Port. Connect an external detector (such as a mass spectrometer) to the analyzer. It cannot be manually controlled. | |||
|
|
Closed Valve. Yellow indicates a closed valve. When manual control is disabled, closed valves appear white. |
|||
|
CryoCooler. Displays only if installed. The CryoCooler
Blue. Indicates the Dewar is full Red. Indicates the Dewar is empty. When the rate speed is set to 100%, the Dewar displays as red and empty, a warning light illuminates, and the pump will beep continuously. Yellow. Indicates Dewar level is unknown and rate is set to zero (for CryoCooler III installations only). |
|||
|
|
Detector Heat Zone. Set the detector heat zone by clicking anywhere in the orange box. Set the Filament Temperature by clicking the detector rectangle on the right side of the schematic.
|
|||
|
External Trigger. Controls the contact closure used to trigger an external mass spectrometer. | |||
|
Furnace. Controls the sample furnace temperatures and cooling level.
|
|||
|
|
Gas Inlet Valves. The inlet displays the gas mnemonic if one is assigned. Right-click the inlet valve to open or close the valve. |
|||
|
|
Heat Zones. Controls the temperature of various parts of the analyzer. Component heat zones are enclosed with an orange rectangle. Click a heat zone icon to display the current temperature and the target temperature. Right-click the icon, then select Set Temperature to change the heat zone’s target temperature. Using the default temperature (110 °C) for all heat zones is recommended to ensure a cleaner system. The trap heat zone controls the temperature of the trap valve. |
|||
|
|
Mass Flow Controller. Controls the flow of gas into the analyzer. The MFC target and flow rate are displayed. Right-click and select an option:
|
|||
|
|
Open Valve. Green indicates an open valve. |
|||
|
|
Pressure Transducer. Displays the current pressure in the carrier gas path, upstream from the septum but downstream from the Mass Flow Controller. It cannot be manually controlled. | |||
|
|
Rotary Valve.
|
|||
|
|
Sample Furnace. Controls the temperature of the sample. Click to display the sample temperature, the target temperature, the furnace temperature, and the ramp rate. Right-click the icon and select Set Temperature to change the Set Point and/or the Ramp Rate. | |||
|
Sample Temperature. Controls the sample Target temperature.
|
|||
|
Trap Heater. Controls the temperature and ramp rate.
|
|||
|
Valve Temperature. Controls the Target temperature.
|
|||
|
Vapor Flask. Controls the Target temperature.
|
|||
|
Vapor Reflux. Controls the Target temperature.
|
Instrument Schematic Shortcut Menus
| Icon or Symbol | Description |
|---|---|
|
Valve options |
Close. Closes the selected valve. Open. Opens the selected valve. |
|
Temperature control options |
Select Set Temperature to modify furnace or heater settings. |







