Specifications for the AutoChem III

Electrical

Voltage 100/120/230 VAC
Frequency 50-60 Hz
Power 1250 VA
Overvoltage category II
Temperature System
Range -100 °C to 1200 °C with CryoCooler option
Ambient -5 °C to 1200 °C without CryoCooler option
Selection Digitally set, 0.1 °C increments
Ramp Rates

1 °C/minute minimum, 100 °C/minute maximum
Maintains scheduled rate within 5 °C or 1 minute, whichever is larger

Heating rates:
≤ 25 °C/min at temperatures ≤ 1200 °C
≤ 50 °C/min at temperatures ≤ 1000 °C
≤ 100 °C/min at temperatures ≤ 800 °C

Cooling rates:
≤ 2 °C/min at temperatures ≥ ambient + 25 °C
≤ 5 °C/min at temperatures ≥ 75 °C
≤10 °C/min at temperatures ≥ 110 °C
≤ 25 °C/min at temperatures ≥ 200 °C

≤ 50 °C/min at temperatures ≥ 300 °C

≤ 100 °C/min at temperatures ≥ 500 °C

Temperature Ramp Rate ± 1% of target

Wetted Materials

Standard System Stainless steel, gold plated nickel/iron (filaments), Kalrez (most O-ring seals), quartz (sample tubes, thermocouple sheath), quartz wool (supports sample in tube), Inconel- sheathed sample thermocouple (can be placed outside tube or sheathed in quartz), PEEK (Poly Ether Ether Ketone - some ferrules), Teflon tubing (most exhaust lines; TCD exhaust is stainless steel for connection of external devices), copper (gas supply lines and BET delay path), nickel plated brass and steel (gas inlet valves), Polyaryletherketone/PTFE composite (rotary valves seals), Silicone/ Teflon sandwich (septum). Buna-N (gas inlet valve seals and O-rings).
System with Vapor Generator Option As above, plus borosilicate glass vapor generator flask, Kalrez flask gasket.
System with Enhanced Chemical Resistant Option As above, but stainless steel manifolds are protected from corrosion with vapor deposited Silcolloy by SilcoTek, stainless steel tubing, fittings, ferrules and rotary valves are replaced by Hastelloy-C versions, Teflon MFC seals, Kalrez is used for all O-ring seals and the seals inside the mass flow controllers, stainless steel grade 316 gas supply lines, stainless steel gas inlet valves with Kalrez seals.
Gases
Loop H2, CO, O2, N2O, NH2 vapors such as pyridine, water, etc.
Carrier He, Ar, and other gases
Preparation H2, O2, He, Ar, and others
Gas Flow Rate
All Mass Flow Controllers (MFCs)
Manual Control 0 to 100 mL/minute1
Automatic Analysis 1 to 100 mL/minute
Gas Delivery
Inlet Ports 18 (6 each for preparation gas, carrier gas and loop gas)
Temperature Control Internal gas lines and valves heated up to 150 °C
Sample Tube
Type Fused quartz Flow-Thru sample tubes, for use up to 1200 °C, accepts powders and pellets up to 9 mm in diameter
Physical
Height 64 cm (25 in.)
Width 61 cm (24 in.)
Depth 64 cm (25 in.)
Weight 87 kg (192 lb.)
Environment
Temperature

15 °C to 35 °C (59 °F to 95 °F), operating
0 °C to 50 °C (32 °F to 122 °F), non-operating
Maximum rate of change of 2 °C per hour

Humidity 20% to 80% relative, non-condensing
Indoor or Outdoor use

Indoor only (not suitable for wet locations)
Altitude: 2000 m max (6500 ft)
Pollution degree of the intended environment: 2

Location Instrument should be located in a dust-free, vibration free environment, away from exposure to direct sunlight and direct air drafts.
Degree of Ingress
Protection
IPX0

Computer Requirements

Operating System

Windows 10 or higher operating system is required.
Desktop Installation Required

The application should not be installed on a network drive with shared access.

Multiple users cannot operate the application at the same time.

 

Ensure the "Sleep" setting on the desktop is set to "Never" to avoid interruption while running an analysis. If this occurs, the application loses network connectivity with the instrument and a communications error will be reported. A restart of the Windows application may be required if automatic reconnection is not successful.

10 Base T or 100 Base T Ethernet Port If the computer is to be connected to a network, two Ethernet ports are required. If more than one Ethernet-based unit is connected to the same computer, an Ethernet switch will also be required.
Read/Write Permissions All application users will need Read/Write permission to all directories and subdirectories where the application is installed.
Drives USB port

Due to continuous improvements, specifications are subject to change without notice.