Specifications for the AutoChem III
| Voltage | 100/120/230 VAC |
| Frequency | 50-60 Hz |
| Power | 1250 VA |
| Overvoltage category | II |
| Range | -100 °C to 1200 °C with CryoCooler option
Ambient -5 °C to 1200 °C without CryoCooler option |
| Selection | Digitally set, 0.1 °C increments |
| Ramp Rates |
1 °C/minute minimum, 100 °C/minute maximum Heating rates: Cooling rates: ≤ 50 °C/min at temperatures ≥ 300 °C ≤ 100 °C/min at temperatures ≥ 500 °C |
| Temperature Ramp Rate | ± 1% of target |
| Standard System | Stainless steel, gold plated nickel/iron (filaments), Kalrez (most O-ring seals), quartz (sample tubes, thermocouple sheath), quartz wool (supports sample in tube), Inconel- sheathed sample thermocouple (can be placed outside tube or sheathed in quartz), PEEK (Poly Ether Ether Ketone - some ferrules), Teflon tubing (most exhaust lines; TCD exhaust is stainless steel for connection of external devices), copper (gas supply lines and BET delay path), nickel plated brass and steel (gas inlet valves), Polyaryletherketone/PTFE composite (rotary valves seals), Silicone/ Teflon sandwich (septum). Buna-N (gas inlet valve seals and O-rings). |
| System with Vapor Generator Option | As above, plus borosilicate glass vapor generator flask, Kalrez flask gasket. |
| System with Enhanced Chemical Resistant Option | As above, but stainless steel manifolds are protected from corrosion with vapor deposited Silcolloy by SilcoTek, stainless steel tubing, fittings, ferrules and rotary valves are replaced by Hastelloy-C versions, Teflon MFC seals, Kalrez is used for all O-ring seals and the seals inside the mass flow controllers, stainless steel grade 316 gas supply lines, stainless steel gas inlet valves with Kalrez seals. |
| Loop | H2, CO, O2, N2O, NH2 vapors such as pyridine, water, etc. |
| Carrier | He, Ar, and other gases |
| Preparation | H2, O2, He, Ar, and others |
| All Mass Flow Controllers (MFCs) | |
| Manual Control | 0 to 100 mL/minute1 |
| Automatic Analysis | 1 to 100 mL/minute |
| Inlet Ports | 18 (6 each for preparation gas, carrier gas and loop gas) |
| Temperature Control | Internal gas lines and valves heated up to 150 °C |
| Type | Fused quartz Flow-Thru sample tubes, for use up to 1200 °C, accepts powders and pellets up to 9 mm in diameter |
| Height | 64 cm (25 in.) |
| Width | 61 cm (24 in.) |
| Depth | 64 cm (25 in.) |
| Weight | 87 kg (192 lb.) |
| Temperature |
15 °C to 35 °C (59 °F to 95 °F), operating |
| Humidity | 20% to 80% relative, non-condensing |
| Indoor or Outdoor use |
Indoor only (not suitable for wet locations) |
| Location | Instrument should be located in a dust-free, vibration free environment, away from exposure to direct sunlight and direct air drafts. |
| Degree of Ingress Protection |
IPX0 |
|
Operating System |
Windows 10 or higher operating system is required. | ||
| Desktop Installation Required |
The application should not be installed on a network drive with shared access. Multiple users cannot operate the application at the same time.
|
||
| 10 Base T or 100 Base T Ethernet Port | If the computer is to be connected to a network, two Ethernet ports are required. If more than one Ethernet-based unit is connected to the same computer, an Ethernet switch will also be required. | ||
| Read/Write Permissions | All application users will need Read/Write permission to all directories and subdirectories where the application is installed. | ||
| Drives | USB port |
Due to continuous improvements, specifications are subject to change without notice.