Preventive Maintenance
Perform the following preventive maintenance procedures to keep the analyzer operating at peak performance. Micromeritics also recommends that preventive maintenance procedures and calibration be performed by a Micromeritics Service Representative every 12 months.
| Maintenance Required | Frequency |
|---|---|
| Banana plug | Check every 600 samples or 3 months. Replace as needed. |
| Chamber plug seals | Replace every 600 samples or 3 months. |
| High pressure chambers | Check and clean every 600 samples or 3 months. |
| High pressure fluid | Check every 600 samples or 3 months. |
| High pressure fluid level | Check prior to performing a high pressure analysis. Fill as needed. |
| Hydraulic pump fluid | Check every 6 months. Fill as needed. Change annually. |
| Low pressure ports | Grease every 600 samples or 3 months. |
| Mercury dish | Check daily for spills. Service as needed. |
| Mercury level | Service daily. |
| Mercury vapor filter | Replace annually. |
| Moisture in system | Remove as needed. |
| Vacuum pump exhaust filter | Replace every 600 samples or 3 months. |
| Vacuum pump fluid | Change every 1200 samples or 3 months. |
| Vacuum pump fluid level | Check monthly. Fill as needed. Refer to the vacuum pump user manual for instructions. |
| Valves | Check for leaks and clean as needed. |