Weigh the Assembled Penetrometer with Sample
| To avoid transferring skin oils, it is recommended to wear latex gloves when handling penetrometers. Skin oils may affect results. |
- Weigh the assembled penetrometer with sample using an analytical balance. Do not include the spacer when weighing.
- Record the mass on the Sample Data Worksheet as Sample + Penetrometer mass.
- Subtract Sample mass from Sample + Penetrometer mass. Record on the Sample Data Worksheet as Penetrometer mass.
| The mass of the penetrometer must be determined by this method in order to account for the mass of the sealing grease, which varies with each application. |